In the field of peptide detection, we rarely compromise by simply running a reverse phase chromatography and providing a purity. The reason is simple: the quality of peptides can never be determined by a single number.
Taking MOTS-c (CAS 1627580-64-6) as an example. The sample submitted by the customer for testing is a 16 peptide with the sequence Met Arg Trp Gln Glu Met Gly Tyr Ile Phe Tyr Pro Arg Lys Leu Arg, belonging to the class of mitochondrial derived peptides. There are several places hidden in this sequence that analysts must pay extra attention to.
Firstly, alkaline residues. The entire peptide contains 4 arginine and 1 lysine, with a high net charge and an alkaline isoelectric point. This type of peptide remains weak on a conventional C18 reverse phase column, and its peak shape is prone to tailing. If the mobile phase system is not adjusted properly, the main peak and impurities will be squeezed together, and the purity value will lose its significance.
Secondly, it contains two methionine residues. Sulfide side chains are easily oxidized to sulfoxides, which may result in the formation of oxidation impurity peaks during storage and waiting for injection of the sample. On the spectrum, it appears as a small peak adjacent to the main peak, which needs to be separated in terms of method and the stability of the automatic sampler needs to be examined.
Thirdly, the presence of counterions. The synthesized peptides purified by the trifluoroacetic acid system often exist in the form of trifluoroacetate salts, and the counterions themselves occupy a part of the mass. Therefore, "chromatographic purity" and "peptide content" are not the same concept - a sample with a purity of 98% may have a peptide content of only 70-80%.
Faced with such samples, our reports are usually supported by several methods: reverse phase high performance liquid chromatography (UV-HPLC) area normalization method provides chromatographic purity and related substances; High resolution mass spectrometry is used to determine precise molecular weight and confirm molecular composition; Amino acid composition analysis is based on theoretical composition and proportion verification; Determination of residual counter ions such as trifluoroacetate by ion chromatography; Combined with factors such as water content, peptide content, and endotoxin (horseshoe crab reagent method), provide a complete answer to whether this peptide can be used.

These methods are not temporary constructions, but routine configurations that have been accumulated in the laboratory through years of research on peptide quality.
The laboratory is open to customers, which is our consistent practice. You can be present throughout the entire process from sample reception, pre-processing, injection to data processing: how to dissolve the sample, how to set the gradient, how to read the spectrum, and how to identify abnormal peaks. You can stand next to the instrument and watch together. When encountering situations such as impurities and main peaks flowing out together, or spectra not matching automatically, analysts will adjust their methods and re collect data on the spot, rather than just giving you an incomprehensible report. The data is traceable and verifiable throughout the entire process from generation to report release.
The laboratory's testing team is composed of doctors and professors, who have been engaged in the research of analysis methods for peptide and protein samples for a long time. The most difficult part of peptide detection is often not sample testing, but spectrum analysis - when the sequence contains non natural residues, fatty acid side chain modifications, or complex disulfide bond connections, the software's automatic matching often gives false positives, relying on the analyst's understanding of spectrum analysis and peptide chemistry. We are willing to sit down and discuss this type of issue with the client, and we are also willing to explain the process of judgment to you.